Automatic calibration
Audible and visible alarms
Three working mode: rate/time/volume
With ultrasonic sensor for drop-counting
Lcd, night operable
English or other language display
Specifications
Drop detector: outside caliper rotary detector. (patented technology)
Control way:
Drop counting and adjusting the delivery rate automatically
Pump type: peristaltic finger
Applicable iv sets:
Compatible with iv sets of any standard (including lucifugal iv sets)
Accuracy:
ín25drop/min: ía2%,
>25drop/min: ía3% (with the drop counting control mode)
ía5% (with the rate control mode after the correct calibration)
Delivery rate range:
1í½400drop/min(1drop/min step)
1.0ml/hí½1200ml/h (1.0ml/hí½99.9ml/h, 0.1ml/h step; >99.9ml/h, 1ml/h step)
Delivery time range: 1miní½1999min (1 min step)
Volume limit range: 0í½9999ml (1 ml step)
Total volume range: 1í½9999ml (1 ml step)
Rapid injection: 600ml/h
Bolus: bolus rate 10-1200ml/hr, bolus volume: max of 100ml
Kvo rate:
When the delivery rate ín10ml/h, kvo is 1ml/h. Otherwise kvo is 3ml/h.
Alarms:
Air-in-line, infusion complete when limited volume has been set, empty container, occlusion, door is open, waiting time over, no drop been detected, operational errors, low battery, other device failures.
Power supply:
Ac power supply: ac 100 to 240vía10%ú¼50/60hzú+
Internal li battery: 11.1v, 2000mah. 6 hours of continuous operation on a new fully charged battery with the rate of 20 drops/min.
Power consumption: ín25va
Safety level: classó±, type cf
Operating conditions: ambient temperature: 5 to 40íµíírelative humidity: 20 to 90%
Storage conditions: ambient temperature: -10 to 55íµíírelative humidity: 10 to 90%
Accessories: ac power line, operational manual
Dimensions: 130mm(l)íß 200mm(w) íß 230mm(h)
Net weight: approximately. 2.0kg