Sterile infusion set for single use
Model: infusion set: GB8368 IS-V
infusion set: GB8368 IS-NV
Product structure: plastic protector for spike, spike, air-inlet, soft tube, drip chamber, flow regulator, fluid filter, and injection site for adding fluid drugâ?¦etc.
Product instruction: The products are made by medical polymer materials. Sterilized by EO gas; sterile, non-toxic, pyrogen free. The products are made by medical soft materials surface is sound and transparent.
Technical indication: in accordance with GB8368 and ISO8536-4.
Scope for use: The products are used for intravenous gravity infusion.
3.5 inch LCD screen for clear vision within 5 meters
Different types of indicator
One-button night mode design for patientâ??s more comfortable rest at night
User friendly numeric setting key
Pull & Push hook design for easier door open and close
Anti-slip pump base support
Finger-tip peristaltic pump with down-stream free-flow protect
Automatic calibration
Audible and visible alarms
Three working mode: rate/time/volume
With ultrasonic sensor for drop-counting
Lcd, night operable
English or other language display
Specifications
Drop detector: outside caliper rotary detector. (patented technology)
Control way:
Drop counting and adjusting the delivery rate automatically
Pump type: peristaltic finger
Applicable iv sets:
Compatible with iv sets of any standard (including lucifugal iv sets)
Accuracy:
ín25drop/min: ía2%,
>25drop/min: ía3% (with the drop counting control mode)
ía5% (with the rate control mode after the correct calibration)
Delivery rate range:
1í½400drop/min(1drop/min step)
1.0ml/hí½1200ml/h (1.0ml/hí½99.9ml/h, 0.1ml/h step; >99.9ml/h, 1ml/h step)
Delivery time range: 1miní½1999min (1 min step)
Volume limit range: 0í½9999ml (1 ml step)
Total volume range: 1í½9999ml (1 ml step)
Rapid injection: 600ml/h
Bolus: bolus rate 10-1200ml/hr, bolus volume: max of 100ml
Kvo rate:
When the delivery rate ín10ml/h, kvo is 1ml/h. Otherwise kvo is 3ml/h.
Alarms:
Air-in-line, infusion complete when limited volume has been set, empty container, occlusion, door is open, waiting time over, no drop been detected, operational errors, low battery, other device failures.
Power supply:
Ac power supply: ac 100 to 240vía10%ú¼50/60hzú+
Internal li battery: 11.1v, 2000mah. 6 hours of continuous operation on a new fully charged battery with the rate of 20 drops/min.
Power consumption: ín25va
Safety level: classó±, type cf
Operating conditions: ambient temperature: 5 to 40íµíírelative humidity: 20 to 90%
Storage conditions: ambient temperature: -10 to 55íµíírelative humidity: 10 to 90%
Accessories: ac power line, operational manual
Dimensions: 130mm(l)íß 200mm(w) íß 230mm(h)
Net weight: approximately. 2.0kg
Specifications
Controls:
Drop counting and adjusting the delivery rate automatically
Pump type: peristaltic finger
Applicable iv sets:
Various iv sets(including lucifugal iv sets)
Accuracy:
25drop/min:2%, >25drop/min:3% (used in the drop counting control way)
5% (used in the volumetric control way with the correct calibration)
Delivery rate range:
1½400drop/min(1drop/min step)
1.0ml/h½1200ml/h (1.0ml/h½99.9ml/h, 0.1ml/h step; >99.9ml/h, 1ml/h step)
Delivery time range: 1min½1999min (1 min step)
Volume limit range: 0½9999ml (1 ml step)
Total volume range: 1½9999ml (1 ml step)
Purge rate: 600ml/h
Kvo rate:
When the delivery rate 10ml/h, kvo is 1ml/h. Otherwise kvo is 3ml/h.
Alarms:
Air-in-line, infusion complete(a volume limit has been set), empty container, occlusion, door open, over time, no drop sensor, operation error, low battery and other device failures.
Power source:
Ac power supply: ac 100 to 240v10%ú¼50/60hzú+
Internal li battery: 11.5v, 2000mah. Battery life with continuous operation is approx. 8 hours (when using a new, full charged battery and running at 20drop/min with any standard of iv set).
Power consumption: 25va
Shock protection: classó, type bf