15 inch Six Standard parameters: ECG, RESP, NIBP, SPO2, 2_TEMP, PR/HR
1.13 Type Arrhythmic Analysis, Multi-Lead ECG Waveforms Display in Phase, Real time S_T segment analysis, pacemaker detection Drug calculation and titratiotable;
2. Efficient resistance to interference of defibrillator and electrosurgical cautery;
3. SPO2 can testing for 0.1% Weak;
4. RA-LL impedance Respiration;
5. Trend Coexist Display;
6. OxyCRG Dynamic View Display;
7. Bed to Bed view Display;
8. Networking capacity and nurse calling system;
9. Options of printing, IBP, VGA and EtCO2;
10. UP to 4 hours working capacity of built-in rechargeable battery;
11.15" high resolution color TFT LCD display;
12. Large volume of tabular and graphic trends information storage and easy to recall;
13. Anti-ESU, anti- defibrillator;
14. Capture dynamic waveforms.
Technical specification:
ECG
Lead Mode: 5 Leads ( R, L, F, N, C or RA, LA, LL, RL,V)
Lead selection: I, II, III, avR, avL, avF, V,
Waveform: 2ch
Lead mode: 3 Leads (R, L, F or RA, LA, LL)
Lead selection: I, II, III,
Waveform: 1ch
Gain: ‘2.5mm/mV, ‘5.0mm/mV, ‘10mm/mV, ‘20mm/mV, auto
HR and Alarm
Range
Adult: 15 ~ 300 bpm
Neo/Ped: 15 ~ 350 bpm
Accuracy: ± 1% or ± 1bpm, which great
Resolution: 1 bpm
Sensitivity: > 200 (uV P-P)
Differential Input Impedance: > 5 MO
CMRR
Monitor: > 105 dB
Operation: > 105 dB
Diagnosis: > 85 dB
Electrode offset potential: ±300mV
Leakage Current: < 10uA
Baseline Recovery: < 3 S after Defi.
ECG Signal Range: ±8 m V (Vp-p)
Bandwidth
Surgery: 1 ~ 15 Hz
Monitor: 0.5 ~ 35 Hz
Diagnostic: 0.05 ~ 100 Hz
Calibration Signal: 1 (mV p-p), Accuracy: ±5%
ST Segment Monitoring Range: Measure and Alarm -2.0 ~ +2.0 mV
ARR Detecting
Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Alarm: Available
Review: Available
RESPIRATION
Method: Impedance between R-F (RA-LL)
Differential Input Impedance: >2.5 MO
Measuring Impedance Range: 0.3~5.0O
Base line Impedance Range: 0 – 2.5 KO
Bandwidth: 0.3 ~ 2.5 Hz
Resp. Rate
Measuring and Alarm Range
Adult: 0 ~ 120 rpm
Neo/Ped: 0 ~ 150 rpm
Resolution: 1 rpm
Accuracy: ±2 rpm
Apean Alarm: 10 ~ 40 S