Uv1901 / uv1901pc specifications:
Photometric methods: transmittance, absorbance, energy, reflectivity
Wavelength range: 190nm-1100nm
Special bandwidth: 1nm
Stray light: ¡ü0.03%t (220nm, 360nm)
Wavelength accuracy: ¡ü¡à0.3nm
Wavelength reproducibility: ¡ü0.1nm
Photometric range: -4a ~ 4a
Photometric accuracy: ¡ü¡à0.3%t (0-100%t)
¡ü¡à0.002a (0-0.5a)
¡ü¡à0.004a (0.5-1a)
Photometric reproducibility: ¡ü0.001a (0-0.5a)
¡ü0.002a (0.5-1a)
Stability: ¡ü ¡à0.004a/h
Noise: ¡ü ¡à0.003a
Baseline flatness: ¡ü ¡à0.0008a
Power: 100-240v, 50/60hz
Light source: deuterium & tungsten-halogen lamp

Accessories:
Uv1900 series spectrophotometer 1 set
Power cable 1 pc
Operation manual 1 pc
Certificate of quality checking 1 pc
Fuses (2 a) 2 pcs
10mm quartz cuvettes 2 cases
Packing list 2 pcs
Compact disc 1 pc (uv1901pc & uv1902pc only)
Rs-232 cable 1 pc (uv1901pc & uv1902pc only)
Dust cover 1 pc

Software specifications:
1. Such operation as photometric measurement, spectrum measurement, quantitative measurement, kinetic measurement are offered in uvcon windows applications.
2. Single and multi-wavelength quantitative measurement can be measured with several kinds of arithmetic calculation methods to meet users¡¯ demand.
3. Curves can be measured and recalled in memory with data-handling of arithmetic calculation smooth, abs to/ from %t conversion and peak pick.
4. The measured data and graphics can be stored on pc/ printed/ output for reports.
Outstanding quality, keeping constant improving


Standard export packing


Recent User Reviews

This user has not received any reviews yet!

Verification Status


 
 
Contact Supplier
Renew

More Items Similiar to: Uv1901 Double Beam Spectrophotometer

May-20-14
 
Specification
Wavelength: 190-1100nm
Band width: 2nm
Wavelength accuracy: +/-0.5nm
Wavelength repeatability: 0.2nm
Wavelength set: auto
Photometric accuracy: +/-0.5%t
Photometric repeatability: 0.2%t
Photometric range: h (500nm)
Noise: +/-0.001a
Stray light: 0.05%t (220nm& 360nm)
Data output: usb
Printer port: parallel port
Display mode: lcd(128*64dots)
Light: deuterium & tungsten lamp
Detector: silicon photodiode
Weight: 14kg

Foam plastic and outside with strong cartons
Apr-23-14
 
Light source lamp: preset germany heraeus deuterium lamp / osram long life halogen tungsten lamp


Uv1800 / uv1800pc specifications:
Photometric methods: transmittance, absorbance, energy, reflectivity
Wavelength range: 190nm-1100nm
Special bandwidth: 2nm
Stray light: ¡ü0.05%t (220nm, 360nm)
Wavelength accuracy: ¡ü¡à0.4nm
Wavelength reproducibility: ¡ü0.2nm
Photometric range: -3a ~ 3a
Photometric accuracy: ¡ü¡à0.3%t (0-100%t)
¡ü¡à0.002a (0-0.5a)
¡ü¡à0.004a (0.5-1a)
Photometric reproducibility: ¡ü0.001a (0-0.5a)
¡ü0.002a (0.5-1a)
Stability: ¡ü ¡à0.001a/h
Noise: ¡ü ¡à0.001a
Baseline flatness: ¡ü ¡à0.0015a
Power: 100-240v, 50/60hz
Light source: deuterium & tungsten-halogen lamp

Accessories:
Uv1800 spectrophotometer 1 set
Power cable 1 pc
Operation manual 1 pc
Certificate of quality checking 1 pc
Fuses (2 a) 2 pcs
10mm quartz cuvettes 2 cases
Packing list 2 pcs
Compact disc 1 pc (uv1800pc only)
Rs-232 cable 1 pc (uv1800pc only)
Dust cover 1 pc

Software specifications:
1. Such operation as photometric measurement, spectrum measurement, quantitative measurement, kinetic measurement are offered in uvcon windows applications.
2. Single-wavelength quantitative measurement can be measured with several kinds of arithmetic calculation methods to meet users¡¯ demand.
3. Curves can be measured and recalled in memory with data-handling of arithmetic calculation smooth, abs to/ from %t conversion and peak pick.
4. The measured data and graphics can be stored on pc/ printed/ output for reports.
Outstanding quality, keeping constant improving


Standard export packing
Apr-23-14
 
¡ü¡à0.004a (0.5-1a)
Photometric reproducibility: ¡ü0.001a (0-0.5a)
¡ü0.002a (0.5-1a)
Stability: ¡ü ¡à0.001a/h
Noise: ¡ü ¡à0.001a
Baseline flatness: / ¡ü ¡à0.002a
Power: 100-240v, 50/60hz
Light source: deuterium & tungsten-halogen lamp

Accessories:
Uv1700 spectrophotometer 1 Set
Power cable 1 Pc
Operation manual 1 Pc
Certificate of quality checking 1 Pc
Fuses (2 a) 2 Pcs
10mm quartz cuvettes 2 Cases
Packing list 3 Pcs
Compact disc 1 Pc (uv1700pc only)
Rs-232 cable 1 Pc (uv1700pc only)
Dust cover 1 pc

Software specifications:
1. Such operation as photometric measurement, spectrum measurement, quantitative measurement, kinetic measurement are offered in uvcon windows applications.
2. Single-wavelength quantitative measurement can be measured with several kinds of arithmetic calculation methods to meet users¡¯ demand.
3. Curves can be measured and recalled in memory with data-handling of arithmetic calculation smooth, abs to/ from %t conversion and peak pick.
4. The measured data and graphics can be stored on pc/ printed/ output for reports.
Outstanding quality, keeping constant improving


Standard export packing
Apr-23-14
 
Introduction:
Uv752 is single beam spectrophotometer that analyzes specimen materials in the ultraviolet and visible spectral d software this product is capable of charting and storing the scan results.


Uv752 specifications:
Model Uv752 752pc
Optical system: single beam, grating 1200 line/mm
Wavelength range: 200nm-1000nm
Wavelength accuracy: ¡à2.0nm
Wavelength reproducibility: 1nm
Wavelength readability: 0.1nm
Special bandwidth: 4nm
Photometric range: 0-125%t, -0.097-2.70a, 0-1999c
Photometric accuracy: ¡à0.5%t
Photometric reproducibility: 0.2%t
Stray light: ¡ü0.5%t (ƛ 220, 340nm)
Stability: ¡ü0.004a/1hr/500nm (after 30min warm up)
Light source: deuterium & tungsten-halogen lamp
Power: 220v 50hz
Display: Lcd(2*20)
Data output: Rs-232c
Analysis software: Without With
Dimensions: 440x370x200mm
Weight: 17kg

Accessories:
1. Uv752 spectrophotometer 1 set
2. Power cable 1 pc
3. Operation manual 1 pc
4. 10mm glass cuvettes 4 cases
5. 10mm quartz cuvettes 2 cases
6. Dust cover 1 pc
7. Printer cable 1 pc



Standard export packing
Aug-17-16
 
Atomic absorption spectrophotometer comes with manual gas control system. It can work as standalone mode for routine analysis and can also be pc controlled.

Salient features:
•Microprocessor based, user friendly, menu driven with printer interface and optional pc attachment through rs232c interface.
•Narrow beam optics fully sealed and quartz over coated mirrors.
•Modes of operation include absorbance, emission and concentration.
•Auto saving of all parameters & auto-bandwidth selection.
•Integral gas control system with panel flow meters.
•Precise knobs for vertical, horizontal and rotational alignment of titanium burner.
•The software has facilities to repeat the result of the sample and data treatment.
•Automatic calculation of percentage and ppm in base material.
•Absorbance mode has the facility to measure sample absorbance for selected number of points and to calculate mean, sd and rsd.
•Optional pc compatibility.
•Microsoft windows® based software for data processing, storage, retrieval, interpretation etc.
•Optional : hydride generator

Software features:
Microsoft windows® based pc operating software.
•Easy user interface.
•Linear / quadratic curve equation to predict the sample concentrations.
•Modes of operation include absorbance, emission, concentration and standard addition.
•Saving facility for the scanned data. Print facility for all the scanned data
•Print facility for all the scanned data in graph and tabular formats.
•Software is 21 cfr part 11 compliant with password method protection, instrument logbook etc

Applications:
•Academia & research
•Soil testing
•Micro nutrients
•Fertilizer
•Bio fertilizer
•Pesticides
Mar-20-21
Supplier From Fuzhou, Fujian, China
 
Achromatic Waveplate

Achromatic waveplate is made from two different substrate materials such as crystal quartz and magnesium fluoride. For the single material,waveplateâ??s working wavelength is very limited because of the dispersion of the material. But Achromatic Waveplate uses two different kinds of materials,with the different dispersion, such waveplate is not sensitive to the wavelength change.

Dual Wavelength Waveplate

Dual wavelength waveplate is a special type of multiple waveplate that provides a specific retardance at two different wavelengths, itâ??s particularly useful when used in conjunction with other polarization sensitive components to separate coaxial laser beams of different wavelength. It is widely used to improve conversion efficiency in solid double frequency laser device.

Polarization Rotator

Due to the rotation activity of natural quartz crystal, it also can be used as polarization rotators so that the plane of input linearly polarized beam will be rotated at special angle which is determined by the thickness of quartz crystal.
Aug-24-17
 
1.CS-820 Colour Spectrophotometer for Textile Advantages
1).Transmittance and reflectance in one instrument
2).Simultaneous SCI/SCE measurement
3).4 Testing Calibers
CS-820 colour spectrophotometer contains 4 kinds of testing calibers and the largest reaches 25.4mm.
4).Double optical path spectrum analysis technology
5).It is with large display screen. So the testing result could be directly and clearly seen without connecting with PC.

2.Colour Spectrophotometer CS-820 Technical Data
Type:CS-820
Measurement Condition:
Reflection: d/8(Diffused lighting, 8 degrees receiving)
Compatible with SCI/SCE Measurement ISO7724/1,DIN5033 Teil7,JIS Z8722 condition C,CIE No.15,ASTM E1164)
Transmission d/0(Diffused lighting, Vertical direction receiving)
Size of integrating sphere:Φ15.2cm, Alvan diffused reflection surface coating
Illumination Light source:CLEDs(entire wavelength balanced LED light source)
Sensor:dual light path sensor array
Wavelength range:400-700nm
Wavelength interval: 10nm
Half spectral width: 5nm
Reflectivity range:0-200%
Reflectivity resolution:0.01%
Observation angle:2°/10°
Measurement light source:A,C,D50,D55,D65,D75,F1,F2,F3,F4,F5,F6,F7,F8,F9,F10,F11,F12
CWF,U30,DLF,NBF,TL83,TL84
Data being displayed: reflectance value and figure, sample's color values, color difference values/graph, pass/fail results, color error tendency, history data color simulation, manual input standard sample, generate measurement report
Measurement time interval:2 seconds
Reflection Apertures:Large Aperture (LAV):2.54cm (=1inch)
Medium Aperture (MAV):1.5cm(=0.6inch)
Small Aperture (SAV):0.75cmx1.0cm(0.3inch x 0.4inch)
Very Small Aperture (VSAV):0.3cmx0.8cm(0.12inch x 0.31inch)
Color space:CIE-L*a*b,L*C*h,CIE-L*u*v,XYZ,Yxy,Transmittance, Hunterlab, Munsell MI,CMYK, RGB,HSB
Color difference formulas:Î?E*ab,Î?E*CH,Î?E*uv,Î?E*cmc(2:1),Î?E*cmc(1:1),Î?E*94,Î?E*00
Î?Eab(Hunter),555color classification
Other colorimetric indices:WI(ASTM E313-10,ASTM E313-73,CIE/ISO, AATCC, Hunter, Taube Berger, Ganz, Stensby),YI(ASTM D1925,ASTM E313-00,ASTM E313-73),Tint(ASTM E313-00,CIE,Ganz),Metamerism index Milm, Stick color fastness, Color fastness, APHA, Pt-Co, Gardner
Repeatability:light splitting reflectivity: standard deviation within 0.08%
color values:E*ab
Jun-30-22
 
Sale Zeiss OPMI ORL Surgical Microscope
Used condition like new, Excellent and ready to work
Come with warranty 12 month
Technical Data:
- It is a universal microscope used in general or other surgery such as dental procedures.
- The microscope has high-quality optics.
- Functions of the microscope: - Electrically controlled: smooth adjustment of the zoom (zoom) and focus (focus)
- Smooth regulation of the light intensity.

Features:
- Smooth adjustment of the total magnification of the optics from 5x to 30x using a grip on the handle
- Smooth focusing with the handle on the handle
- Focal length of the lens: f = 250 mm
- Magnification of the doctor's glasses1: 12.5x
- Magnification of the doctor's glasses2: 12.5x
- Adjustable tilt angle (0-180) of the doctor's view1
- Adjusting the distance between the glasses
- Correction of vision defects at both tubes
- S5 tripod
- Light source: 1 x 100 Watt halogen
- A beam of light transmitted to the optics by means of optical fiber
- A double brake system located at the base of the microscope

Dimensions:
- Height of the microscope: (minimum: 160 cm - maximum: 190 cm)
- The maximum distance from the objective lens to the column: 143 cm
- Basis: 64 x 68 cm
Aug-20-19
 
Optical system
Finite color corrected optical system
Head
Monocular head 30°inclined, with locking system
Binocular head 30°inclined, 360° rotatable with locking system,
right eyepiece tube diopter adjustable, ±5D interpupillary distance 50-75mm
Trinocular head,30°inclined, 360° rotatable,
Interpupillary distance:50-75mm,beam splitting ratio R:T=5:5
Eyepiece
Wide field plan eyepiece PL10X/18mm, reticle, pointer optional
Objective
Achromatic objective (4Xã??10Xï¼?40Xï¼?100X)
Plan achromatic objective (4Xã??10Xã??20Xã??40Xã??100X)
Nosepiece
Reversed quadruple nosepiece
Main body
Low position coaxial coarse and fine adjustment, Coarse adjustment range 13 mm
(double layer stage 7mm), precision of fine adjustment:2um with locking system
0.2W LED illumination (color temperature 5700-6500K),
brightness adjustable. Battery charging system is optional
input 100V-240V , output 5V 1A
Stage
Single stage, size 110mmÃ?120mm with two clips,
single lens condenser N.A=0.65, five hole iris diaphragm plate
Double layer mechanical stage: 115mmX125mm ABBE condenser N.A.=1.25,
with iris diaphragm, with fish eye lens holder
C mount
0.35X/0.5X/0.65X/1X C-mount, focus adjustable