NC 654Performance Specifications
Measuring Strokes : 600(mm) x 500(mm) x 400(mm)
Overall Dimension : 1630(mm) x 1280(mm) x 2380 (mm)
Max part weight : 500 kg
Resolution : 0.1 (mm)
Accuracy according GB/T16857.2-1997 equal ISO 10 360ª
(Volumetric length measuring error) E =2.5 +3.5L/1000 [mm](Volumetric Probing Error) R = 2.4 mm
Environment Specifications
Room required specifications
Average room temperature: 18 - 22 pC
Maximum thermal gradients:1 C/h2 pC/day1 pC/m
Air supply system
(Min. air pressure) : 0.7 MPa
Electrical requirements
- (Voltage) : 220 Vac + 10%
Frequency) : 50 Hz
(Current) : 10 A
Allowed Maximum Floor Vibration (Frequency Range): 0 ~ 19 Hz 19~75 Hz > 75 Hz
Supplier: CMM, coordinate measuring machine, laser scanner.
Services: Upgrade of coordinate measurement machine