This buylead is archived. Your inquiry will be sent directly to the buyer's email inbox, and they may respond at their discretion. However, you will not receive their contact details.
More Items Similiar to: Wanted : Ph Measuring Meters
Please provide a quotation to the following requirement from importer -
Product Name: Portable pH And EC Meters
Specification:
Application and function: Measurement of pH-EC in liquids such as water
Portable pH-EC meter with innovative high-resolution color display
Measurement of pH, mV ORP, conductivity, TDS and salinity
Possibility to select the displayed parameters
Automatic pH calibration with USA and NIST buffers up to 3 points
Adjustable reference temperature and selectable cell constant,
All operations and errors are constantly monitored and shown.
Workable in any conditions, Adjust manually, Simple calibration mode, rugged carrying case
Measuring Range:
Measuring range pH 0.00 ... 14.00 (0.1 - 0.01pH)
Measuring range mV -1000 ... +1000 (1 mV)
Measuring range EC 0.1 μS 200 mS (0.01 μS)
Measuring range TDS 0.01 mg/l 200.0 g/l
Measuring range temperature 0.0 +100.0 °C (0.1 °C)
Certificate and Compliance to Standards:
International Protection Class - IP65 (except eye-piece). Dust-tight and Protected against water jets.
Calibration Certificate from Manufacturer
Manuals:
Operators manual and operating instructions in English and Arabic (optional)
Please provide a quotation to the following requirement from the importer -
Product Name: Optical Density Measurement Meters
Specifications:
In-Situ Turbidity Meters For Biomass Monitoring. Suitable For SIP/CIP Cycles.
Model -Trans-4 (V)
Hard Rigid Material
Size: Standard
Packaging Terms : Standard
2
All instrument multifunction meters shall Measure the
following:
The current, the voltage, active power, reactive power, apparent
power, active energy, reactive energy, power factor and
frequency.
3
Accurate measured values with error limits:
Less than 0.5% for the Voltage (V) and the Current (I).
4
Good legibility with high-contrast:
At least 14 mm SCREEN or LED displays.
5 Electrically isolated current inputs.
6 Four quadrant operations.
7
Connection options:
- Single-phase.
- Or three-phase 3 and 4-wires.
- Suitable for balanced or unbalanced load.
8
Input voltage:
- Phase-to-phase Vp: up to 240 V.
- Programmable (primary and secondary).
9
Nominal input current:
1 A (primary should be programmable)
10 Dimensions:
At least (96 X 96 X ---) mm.
11 Auxiliary supply:
110 ± 10% V DC.
12 Multiple displays:
At least three lines.
13
Outputs:
- Four (4) Analog outputs for (MW, MVAR, V, A) shall be
choosable and programmable using the software tool .
- Each output can be programmed for any quantity.
Contact me/send me your offer via ramzi.v.saba@gmail.com
51
Schedule A - Continued
TECHNICAL PARTICULARS AND GUARANTEES
Item
No
Description
(DATA PARAMETERS REQUIREMENT)
Offered
14
Programmable inputs primary and secondary for C.T, V.T and
the Four required Analogue outputs programmable to any value
inside the required full range:
(-20 mA DC to +20 mA DC).
15 Response time: = 300 ms.
16
The continuous overload rating of the device shall:
1- Shall not be less than 1.3 times rated voltage.
2- Shall not be less than 2 times the rated current.
17
- Programming kit (configuration software + communication
cables) Via Rs 485 port or usb port.
- An additional USB programming port shall be applied.
18
The connection between the device and the RTU in the
substation is done through Modbus/TCP on Ethernet and
Modbus ASCII and Modbus RTU modes with the following
frame structure on RS-485 interface:
- ASCII framing (7-bit ASCII characters).
- RTU framing (8-bit characters)
19
The RS-485 serial communication must have a configurable
range of communication speeds:
- 4800 Bits/seconds .
- 9600 Bits/seconds .
- 19200 Bits/seconds.
With any of the following transmission settings for the RS-485
interface:
- even parity, 1 stop bit
- odd parity, 1 stop bit
- none parity, 1 stop bit
An additional USB programming port shall be applied.